LEV testing - fume cupboards

Face velocity at a fume cupboard: what's adequate?

A fume cupboard works by drawing air in across its opening, carrying fumes away from the user and out. The speed of that inward air - the face velocity - is a key measure of whether the cupboard is containing the fumes. Guidance commonly puts an adequate face velocity around 0.5 metres per second, with a lower action level below which it shouldn't be used - but face velocity alone isn't the whole story. Here is what adequate means, and why containment matters too. This is general information, not a specification - follow the standards and your tester's guidance for your cupboard.

Face velocity
The inward air speed at the opening
Around 0.5 m/s
Commonly cited as adequate
Not the whole story
Containment matters too
Established 2001UK-wideFully insuredCertificated and photographed

The short answer

A fume cupboard's face velocity is the speed of the air being drawn inward across its opening (its face, or sash opening) - and it's a key measure of whether the cupboard is working, because that inward flow is what carries fumes away from the user and stops them escaping into the room; as to what's adequate, guidance commonly puts an adequate face velocity around 0.5 metres per second (a range of roughly 0.3 to 0.5 m/s is often cited for standard work), typically measured with the sash (the front opening) at a defined height, and there's usually an action level - a figure such as around 0.45 m/s - below which the cupboard should be taken out of use until it's put right (because below that, containment can't be relied on); but - and this is important - face velocity alone isn't the whole story: it's a useful, measurable indicator, but the real question is containment (does the cupboard actually keep the fumes inside and away from the user), and a cupboard can have an apparently adequate average face velocity yet still fail to contain properly if the airflow is uneven, turbulent, or disrupted (by too high a velocity causing turbulence, by the user's movements, by items blocking the airflow, or by draughts and room air currents), so face velocity is measured alongside checks of the airflow pattern and containment (for example smoke tests showing the air actually flows in and carries away, and sometimes containment testing); so an adequate fume cupboard has a face velocity in the right range and genuinely contains the fumes - the velocity is a key number, but it's judged together with the airflow behaviour and containment, by a competent tester against the relevant standards; so adequate means the right face velocity plus real containment; so it's about the inward air speed and whether fumes are actually kept in; this is general information, not a specification

A fume cupboard (or fume hood) protects the user by drawing air inward across its opening, so that fumes from the work inside are carried away from the person and out through the extraction, rather than escaping into the room. The speed of that inward air - the face velocity - is one of the key things measured to judge whether the cupboard is working. So what counts as an adequate face velocity, and is it the whole story? Here's the picture. This is general information, not a specification - follow the relevant standards and your tester's guidance for your specific cupboard. What face velocity is. Face velocity is the speed of the air being drawn inward across the fume cupboard's opening - its face, or the sash opening (the sash being the movable front, usually adjustable in height). As air is extracted from the cupboard, room air is drawn in across the opening to replace it, and the speed of that inward flow is the face velocity, measured in metres per second (m/s). It matters because that inward flow is what keeps fumes inside the cupboard and away from the user: if air is flowing inward across the whole opening at an adequate speed, fumes generated inside are carried inward and away, not out into the user's face. So face velocity is a direct measure of the containing airflow. What's adequate. Guidance commonly puts an adequate face velocity around 0.5 metres per second for standard work with hazardous substances (a range of roughly 0.3 to 0.5 m/s is often cited, depending on the situation and the standard). It's typically measured with the sash at a defined height (a set opening), because the velocity changes with how far the sash is open - so the measurement is standardised to a particular sash position. And there's usually an action level - a figure such as around 0.45 m/s in some guidance - below which the cupboard should be taken out of use until it's put right, because below that the containment can't be relied on. So there's a target range (around 0.5 m/s) and a floor (an action level) that shouldn't be breached. These are general figures - the precise values, sash height and method come from the relevant standards and should be applied by a competent tester; this page gives the general picture, not the exact specification for your cupboard. Why it's not the whole story. Here's the important caveat: face velocity alone isn't the whole story. It's a useful, measurable indicator, but the real question is containment - does the cupboard actually keep the fumes inside and away from the user? A cupboard can have an apparently adequate average face velocity yet still fail to contain properly, for several reasons. Uneven airflow: the average velocity might be fine, but the flow could be uneven across the opening (strong in places, weak in others), leaving weak spots where fumes can escape. Turbulence: too high a face velocity can actually cause turbulence and eddies at the opening that pull fumes out, so more isn't simply better. Disruption: the user's movements, items placed near the opening blocking the airflow, or draughts and room air currents (from doors, people passing, other ventilation) can disrupt the inward flow and let fumes escape, even if the measured velocity looks right. So a single average face-velocity number doesn't guarantee containment. How it's actually judged. Because of this, a proper fume cupboard test measures face velocity alongside checks of the airflow pattern and containment. That includes: measuring the face velocity at multiple points across the opening (to check it's adequate and reasonably even, not just adequate on average); observing the airflow pattern, often with a smoke test (releasing smoke to see whether the air actually flows inward and carries away across the whole opening, with no roll-back or escape); and sometimes containment testing (a more formal check of whether a tracer released inside stays contained). So the tester judges the cupboard on whether it genuinely contains the fumes - with face velocity as a key measurement within that, not the sole criterion. This is why fume cupboard testing is more than a single velocity reading (covered in the lab-fume-cupboards-testing-beyond-the-standard-lev-check page). Using it well. So face velocity is a key number to know and check - if it's below the action level, the cupboard shouldn't be relied on and needs attention; if it's in the right range, that's necessary but not by itself sufficient. Good practice also means using the cupboard correctly (working with the sash at the right height, not blocking the airflow, keeping the work back from the opening, avoiding rapid movements) and keeping the room environment stable (minimising draughts and disruptions), so the containment the velocity should provide isn't undermined in use. And it means having the cupboard properly tested (face velocity plus airflow and containment) at the appropriate intervals by a competent tester. The takeaway. So an adequate fume cupboard face velocity is commonly around 0.5 m/s (with a range often cited and an action level below which it shouldn't be used), measured at a defined sash height. But face velocity alone isn't the whole story - the real measure is whether the cupboard actually contains the fumes, which depends on even, stable, undisrupted airflow as well as the velocity. So it's judged together with the airflow pattern and containment, by a competent tester against the relevant standards. This is general information, not a specification. This is general information.

Key points

The short version

  • Face velocity is the speed of air drawn in across the fume cupboard's opening.
  • It's a key measure of whether the cupboard contains fumes away from the user.
  • Guidance commonly puts adequate face velocity around 0.5 m/s (a 0.3-0.5 range is often cited).
  • There's usually an action level below which the cupboard should be taken out of use.
  • Face velocity alone isn't the whole story - containment is what ultimately matters.

What face velocity is

The inward air speed at the opening

Face velocity is the speed of the air being drawn inward across the fume cupboard's opening - its face, or the sash opening (the sash being the movable front, usually adjustable in height). As air is extracted from the cupboard, room air is drawn in across the opening to replace it, and the speed of that inward flow is the face velocity, measured in metres per second (m/s).

It matters because that inward flow is what keeps fumes inside the cupboard and away from the user. If air is flowing inward across the whole opening at an adequate speed, fumes generated inside the cupboard are carried inward and away (into the extraction), not out into the user's breathing zone. If the inward flow is too weak, fumes can drift or roll out of the opening towards the user. So face velocity is a direct, measurable indicator of the containing airflow - which is why it's a headline figure in judging a fume cupboard. What counts as adequate is next. So what face velocity is - the inward air speed at the opening - is the basic measure: the speed of air being drawn in to carry fumes away from the user. So it's the speed of air pulled in at the front. This is general information. This is general information.

What's adequate

Commonly around 0.5 m/s, with an action level

Guidance commonly puts an adequate face velocity around 0.5 metres per second for standard work with hazardous substances (a range of roughly 0.3 to 0.5 m/s is often cited, depending on the situation and the standard). It's typically measured with the sash at a defined height (a set opening), because the velocity changes with how far the sash is open - so the measurement is standardised to a particular sash position for consistency.

And there's usually an action level - a figure such as around 0.45 m/s in some guidance - below which the cupboard should be taken out of use until it's put right, because below that the containment can't be relied on. So there's a target (around 0.5 m/s) and a floor (an action level) that shouldn't be breached. These are general figures: the precise values, the sash height, and the measurement method come from the relevant standards, and should be applied by a competent tester for your specific cupboard - this page gives the general picture, not the exact specification. Note too that higher isn't simply better (too high causes its own problems, covered next). So what's adequate - commonly around 0.5 m/s, with an action level - gives the ballpark: aim for the right range, don't drop below the action level, and get the exact figures from the standards. Why the number isn't the whole story is next. So adequate is around 0.5 m/s, above the action level. This is general information, not a specification. This is general information.

Why the number isn't everything

Containment is what really matters

Here's the important caveat: face velocity alone isn't the whole story. It's a useful, measurable indicator, but the real question is containment - does the cupboard actually keep the fumes inside and away from the user? A cupboard can have an apparently adequate average face velocity yet still fail to contain properly, for several reasons.

Uneven airflow: the average velocity might be fine, but the flow could be uneven across the opening (strong in places, weak in others), leaving weak spots where fumes escape. Turbulence: too high a face velocity can cause turbulence and eddies at the opening that actually pull fumes out - so more isn't simply better. Disruption: the user's movements, items placed near the opening blocking the airflow, or draughts and room air currents (from doors, passing people, other ventilation) can disrupt the inward flow and let fumes escape, even if the measured velocity looks right. So a single average face-velocity number doesn't guarantee containment - the airflow has to be adequate, even, stable and undisrupted to actually contain. So why the number isn't everything - containment is what really matters - is the key caveat: face velocity is necessary but not sufficient; genuine containment is the real test. How that's judged is next. So the velocity number alone doesn't prove containment. This is general information. This is general information.

How it's actually judged

Velocity, airflow pattern and containment together

Because of this, a proper fume cupboard test measures face velocity alongside checks of the airflow pattern and containment. That includes: measuring the face velocity at multiple points across the opening (to check it's adequate and reasonably even, not just adequate on average); observing the airflow pattern, often with a smoke test (releasing smoke to see whether the air actually flows inward and carries away across the whole opening, with no roll-back or escape towards the user); and sometimes containment testing (a more formal check of whether a tracer released inside the cupboard stays contained).

So the tester judges the cupboard on whether it genuinely contains the fumes - with face velocity as a key measurement within that, not the sole criterion. A cupboard passes not just by hitting a velocity number but by demonstrating that it contains: adequate, even velocity, a sound inward airflow pattern, and no escape. This is why fume cupboard testing is more than a single velocity reading (covered in the lab-fume-cupboards-testing-beyond-the-standard-lev-check page) - and why it should be done by a competent tester who assesses the whole picture against the relevant standards. So how it's actually judged - velocity, airflow pattern and containment together - is the proper approach: the face velocity is measured and judged alongside the airflow behaviour and real containment. Using the cupboard well matters too (next). So a good test checks velocity and containment together. This is general information. This is general information.

Using it well

Correct use keeps the containment real

So face velocity is a key number to know and check: if it's below the action level, the cupboard shouldn't be relied on and needs attention; if it's in the right range, that's necessary but not by itself sufficient. Alongside the testing, using the cupboard correctly is what keeps the containment the velocity should provide from being undermined in day-to-day use.

Good practice in use includes: working with the sash at the right height (as tested - the sash position affects the velocity and containment); keeping the work back from the opening (not right at the front edge, where escape is easier); not blocking the airflow with large items or clutter near the opening (which disrupts the inward flow); avoiding rapid movements at the opening (which can pull fumes out); and keeping the room environment stable (minimising draughts, and not siting the cupboard where doors, passing traffic or other ventilation disturb its airflow). These keep the tested containment real in practice. And it means having the cupboard properly tested - face velocity plus airflow and containment - at the appropriate intervals by a competent tester (a fume cupboard is a type of LEV, so it falls under the regime for thorough examination and testing). So using it well - correct use keeps the containment real - completes the picture: a good face velocity, verified by proper testing, and preserved by correct use and a stable environment. So use it correctly and test it properly. This is general information. This is general information.

Questions

Frequently asked questions

What is face velocity at a fume cupboard?

It's the speed of the air drawn inward across the cupboard's opening (its face or sash opening), measured in metres per second; that inward flow carries fumes away from the user and out through the extraction, so face velocity is a key measure of whether the cupboard is containing the fumes. Face velocity at a fume cupboard is the speed of the air being drawn inward across the cupboard's opening - its face, or the sash opening (the sash being the movable front, usually adjustable in height). As air is extracted from the cupboard, room air is drawn in across the opening to replace it, and the speed of that inward flow is the face velocity, measured in metres per second (m/s). It matters because that inward flow is what keeps fumes inside the cupboard and away from the user: if air flows inward across the whole opening at an adequate speed, fumes generated inside are carried inward and away into the extraction, rather than escaping towards the user's breathing zone. If the inward flow is too weak, fumes can drift or roll out of the opening towards the user. So face velocity is a direct, measurable indicator of the containing airflow - which is why it's a headline figure in judging whether a fume cupboard is working. It's measured at a defined sash height, because the velocity changes with how far the sash is open. So it's the inward air speed at the opening that carries fumes away. So it's the speed of air pulled in across the front. This is general information. This is general information.

What face velocity is adequate for a fume cupboard?

Guidance commonly puts it around 0.5 m/s for standard work (a 0.3-0.5 m/s range is often cited), measured at a defined sash height, with an action level (such as around 0.45 m/s) below which the cupboard should be taken out of use; the exact figures come from the relevant standards and a competent tester. Guidance commonly puts an adequate fume cupboard face velocity around 0.5 metres per second for standard work with hazardous substances - with a range of roughly 0.3 to 0.5 m/s often cited, depending on the situation and the standard. It's typically measured with the sash at a defined height (a set opening), because the velocity changes with how far the sash is open, so the measurement is standardised to a particular sash position. There's usually also an action level - a figure such as around 0.45 m/s in some guidance - below which the cupboard should be taken out of use until it's put right, because below that containment can't be relied on. So there's a target (around 0.5 m/s) and a floor (the action level) that shouldn't be breached. These are general figures - the precise values, sash height and method come from the relevant standards and should be applied by a competent tester for your specific cupboard. And note that higher isn't simply better, because too high a velocity causes turbulence. So it's commonly around 0.5 m/s, above the action level, per the standards. So around 0.5 m/s, with the exact figures from the standards. This is general information, not a specification. This is general information.

Is a higher face velocity always better?

No - too high a face velocity can cause turbulence and eddies at the opening that actually pull fumes out towards the user, so more isn't simply better; there's a right range, and both too low (weak containment) and too high (turbulence) are problems - which is why the velocity is judged against a range, not maximised. No - a higher face velocity is not always better. While too low a face velocity means the inward flow is too weak to contain the fumes (they can drift or roll out towards the user), too high a face velocity causes its own problem: turbulence and eddies at the opening. Air rushing in too fast creates turbulent flow and swirls at the cupboard's opening, and those eddies can actually pull fumes out of the cupboard towards the user - the opposite of what's wanted. So more velocity isn't simply better; beyond a point it degrades containment rather than improving it. This is why an adequate face velocity is defined as a range (commonly around 0.5 m/s), not just a minimum to exceed as far as possible - both too low (weak containment) and too high (turbulence) are problems. The aim is a velocity in the right range, giving a smooth, adequate inward flow that contains the fumes without creating turbulence. So the velocity should be in the correct range, not maximised. So no - too high causes turbulence; aim for the right range. So no; there's a right range, not the higher the better. This is general information. This is general information.

Does an adequate face velocity mean the fume cupboard is safe?

Not by itself - face velocity is a key measure but not the whole story; a cupboard can have an adequate average velocity yet fail to contain if the airflow is uneven, turbulent, or disrupted by movements, clutter or draughts, so containment (whether fumes are actually kept in) is the real test, checked alongside the velocity. Not by itself - an adequate face velocity is necessary but not sufficient. Face velocity is a key, measurable indicator, but the real question is containment: does the cupboard actually keep the fumes inside and away from the user? A cupboard can have an apparently adequate average face velocity yet still fail to contain properly, for several reasons. The airflow could be uneven across the opening (adequate on average but weak in places, leaving spots where fumes escape). Too high a velocity can cause turbulence that pulls fumes out. And disruption - the user's movements, items blocking the airflow near the opening, or draughts and room air currents - can disrupt the inward flow and let fumes escape even if the measured velocity looks right. So a single average face-velocity number doesn't guarantee containment. That's why proper testing checks the airflow pattern (often with a smoke test) and containment alongside the velocity, and why correct use and a stable environment matter. So an adequate velocity is a good sign but not a guarantee - real containment is the test. So not alone - containment, checked alongside, is the real measure. So not by itself; containment is what really matters. This is general information. This is general information.

How is a fume cupboard properly tested?

By measuring face velocity at several points across the opening (adequate and even), observing the airflow pattern (often a smoke test to see the air flows in and carries away with no escape), and sometimes formal containment testing - judged together by a competent tester against the relevant standards, at appropriate intervals. A fume cupboard is properly tested by measuring face velocity alongside checks of the airflow pattern and containment - not by a single velocity reading. A proper test includes: measuring the face velocity at multiple points across the opening (to check it's adequate and reasonably even, not just adequate on average, at the defined sash height); observing the airflow pattern, often with a smoke test (releasing smoke to see whether the air actually flows inward and carries away across the whole opening, with no roll-back or escape towards the user); and sometimes containment testing (a more formal check of whether a tracer released inside stays contained). So the tester judges the cupboard on whether it genuinely contains the fumes - with face velocity a key measurement within that, not the sole criterion. A fume cupboard is a type of LEV, so it falls under the regime for thorough examination and testing (typically at least every fourteen months), and should be tested by a competent tester who assesses the whole picture against the relevant standards. So it's tested on velocity, airflow pattern and containment together. So by checking velocity, airflow and containment, competently. This is general information. This is general information.

What can I do to keep my fume cupboard working well?

Use it correctly - work with the sash at the tested height, keep work back from the opening, don't block the airflow with clutter, avoid rapid movements, and minimise draughts around it - and have it properly tested (velocity, airflow and containment) at the right intervals by a competent tester. To keep your fume cupboard working well, combine correct use with proper testing. Correct use keeps the containment real in day-to-day work: work with the sash at the right (tested) height, as the sash position affects the velocity and containment; keep the work back from the opening rather than right at the front edge (where escape is easier); don't block the airflow with large items or clutter near the opening (which disrupts the inward flow); avoid rapid movements at the opening (which can pull fumes out); and keep the room environment stable, minimising draughts and not letting doors, passing traffic or other ventilation disturb the cupboard's airflow. These preserve the containment the face velocity should provide. Alongside that, have the cupboard properly tested at the appropriate intervals by a competent tester - face velocity measured across the opening, plus the airflow pattern (smoke test) and containment - since a fume cupboard is a type of LEV subject to thorough examination and testing. So it's correct daily use plus proper periodic testing. So use it correctly and keep it properly tested. So use it well and test it regularly. This is general information. This is general information.

More guides

Read next

Test your fume cupboard on containment

A fume cupboard's face velocity is a key measure, but real protection depends on whether it actually contains the fumes; we test fume cupboards on face velocity, airflow pattern and containment together, against the relevant standards, so you know the cupboard genuinely protects the user - not just that a number is met. Ask us to test your fume cupboards. This is general information.